Cat.no. DT-0500-0C

  • The device is supplied with tap water.
  • Degrees of water purification:
    • filtration on sediment filters:
      cascade filtration on 20µm and 5µm high efficiency sediment filters,
    • filtration on carbon filters:
      filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,
    • softening process (option):
      automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,
    • reverse osmosis station:
      efficiency: 400 – 600 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,
    • UV lamp (option).
  • Automatic and maintenance-free operation of the device.
  • Retention rate is 96-99%.
  • Water intake point – third * purity class according to ISO 3696: 1999.
  • Tank for storing purified water (capacity to choose).
  • Automatic system shutdown when the tank is full or the water intake is closed.
  • Can be connected to a dishwasher, autoclave, analyzer, etc.
  • Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.
  • Control water intake points.
  • Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).
  • Forced flushing of membranes (service).
  • System designed for cold water supply: 5-40 ° C.
  • Possibility of self service by the User (without having to call the service).
  • Power supply: 230V / 50Hz.
  • Stainless steel frame.

Functions monitoring system operation:

  • The device is equipped with a microprocessor control and measuring system having:
    • color graphic display with Touch Panel function,
    • conductivity meter measuring the conductivity and temperature of tap water,
    • conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,
    • conductivity measurement in µS / cm or MOhm units,
    • automatic temperature compensation,
    • continuous control and preview of the degree of retention (degree of retention) of RO membranes,
    • clock displaying date and time,
    • alarm informing about mechanical and carbon filter replacement,
    • alarm informing about replacement of the RO module,
    • alarm informing about the UV lamp radiator replacement (option),
    • information about the tank filling level on the device display,
    • preview of service dates,
    • menu in Polish, English, Russian or Spain on display,
    • built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,
    • built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,
    • computer program enabling individual settings of alarm thresholds and data archiving,
    • built-in manometer.

Functions securing the system operation:

  • Interruption of system operation with:
    • low feed water pressure (no feed water),
    • a full tank / closed water intake point.
  • Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.
  • Ability to stop the system when any alarm occurs.
  • System autostart capability.
  • Preview of monitoring messages / alarms.
  • Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.
    * depends on the quality of the feed water, the degree of retention is 96-99%

Additional equipment / services:

  • tank for storing purified water:
    • pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,
    • pressureless – capacity to be agreed (on request),
  • compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),
  • automatic softening station,
  • tank housing made of stainless acid-resistant steel,
  • full DQ, IQ, OQ, PQ qualification procedure with documentation,
  • cooperation with building management systems (BMS).

Required connections at the installation site:

  • cold tap water connection ¾” or 1″,
  • drain to sewage system (sewage grate),
  • 230V socket.